IAQ-CALC
Indoor Air Quality Meters


Models 8732, 8760, 8762
TSI's IAQ-Calc™ Meters are exceptional instruments for investigating and monitoring Indoor Air Quality (IAQ). The Model 8732 is a cost-effective meter for CO2 measurements. The Models 8760 and 8762 simultaneously measure and data log multiple parameters to monitor indoor air quality conditions. The Model 8760 measures CO2, temperature, humidity, and calculates dew point, wet bulb temperature, absolute humidity, humidity ratio and % outside air. The Model 8762 adds CO.


Features and Benefits

ALL MODELS

  • Low-drift NDIR CO2 sensor ensures stable, accurate readings
  • Sampling function records multiple point measurements
  • Optional portable printer provides hard copy documentation

MODELS 8760 and 8762

  • Temperature and relative humidity measurements help determine thermal comfort levels
  • Calculates % outside air from either CO2 or temperature readings
  • Directly calculates dew point, wet bulb temperature, absolute humidity and humidity ratio — no psychrometric chart needed
  • Electrochemical sensor measures CO at outdoor air intakes, HVAC systems and parking garages (Model 8762 only)
  • Two-line display shows multiple parameters simultaneously
  • Optional detachable secondary temperature sensor measures temperature at heat sources and investigates thermal gradients

Data Logging Capabilities

MODEL 8760 and 8762

  • Data logging allows user to log 14,000 samples with a time and date stamp
  • Simultaneously records all parameters available
  • Continuous data logging enables full-cycle monitoring and trend analysis
  • TSI LOGDAT™ downloading software permits easy transfer of data to a computer
  • LOGDAT Companion Excel® Macro automatically graphs your data
  • Data can be reviewed on-screen, printed or downloaded to a computer for easy report generation
  • Statistics function displays average, maximum and minimum values, and the number of recorded samples

Applications

  • Measure IAQ and comfort conditions in schools, offices and other indoor environments
  • Investigate Sick Building Syndrome and building related illnesses
  • Verify building HVAC system performance
  • Examine building IAQ conditions to optimize worker productivity
  • Comply with regulations and guidelines

IAQ-CALC™ Meter Specifications--Models 8732, 8760, and 8762

CO2:
  Sensor Type:                      Dual-wavelength NDIR (non-dispersive infrared)
  Range:                            0 to 5000 ppm
  Accuracy1:                        ±3.0% of reading or ±50 ppm, whichever is greater  
  Resolution:                       1 ppm
Temperature (Models 8760 and 8762 only) 
  Sensor Type:                      Thermistor
  Range:                            0 to 60°C (32 to 140°F)
  Accuracy:                         ±0.6°C (±1.0°F)  
  Resolution:                       0.1°C (0.1°F)
Optional Secondary Temperature Probe 
(Models 8760 and 8762):
  Sensor Type:                      RTD
  Range:                            -40 to 120°C (-40 to 248°F)
  Accuracy2:                        ±0.6°C (±1.0°F)  
  Resolution:                       0.1°C (0.1°F)
Relative Humidity (Models 8760 and 8762):
  Sensor Type:                      Thin-film capacitive
  Range:                            5% to 95% RH
  Accuracy3:                        ±2.0% RH
  Resolution:                       0.1% RH
Dew Point (Models 8760 and 8762):
  Range:                            -15 to 49°C (5 to 120°F)
  Resolution:                       0.1°C (0.1°F)
Wet Bulb Temperature (Models 8760 and 8762):
  Range:                            5 to 60°C (40 to 140°F) 
  Resolution:                       0.1°C (0.1°F) 
Absolute Humidity (Models 8760 and 8762):
  Range:                            0.24 to 79.0 g moisture/m3 air
                                    (.0000151 to .00491 lbs moisture/ft3 air) 
Humidity Ratio (Models 8760 and 8762):
  Range:                            0.14 to 187 g moisture/kg dry air
                                    (0.00014 to 0.187 lbs moisture/lb dry air)                          
% Outside Air (Models 8760 and 8762):
  Range:                            0 to 100% 
  Resolution:                       0.1% 
CO (Model 8762):
  Sensor Type:                      Electro-chemical 
  Range:                            0-500 ppm 
  Accuracy:                         ±3.0% of reading or ±3 ppm, whichever is greater
  Resolution:                       0.1 ppm
  Response Time:                    <60 sec to 90% step change
Operating Temperature:
  Models 8732 and 8760:             5 to 70°C (41 to 158°F)
  Model 8762:                       5 to 45°C (41 to 113°F)
Storage Temperature:                -20 to 70°C (-4 to 158°F)
Logging Capability (Models 8760 and 8762):
  Range:                            Up to 14,000 samples and 1,000 test id's
  Intervals:                        5 sec, 10 sec, 15 sec, 20 sec, 30 sec, 60 sec, 
                                    2 min, 5 min, 10 min, 15 min, 20 min, 30 min, 60 min
Meter Dimensions:                   100 mm x 168 mm x 38 mm
                                    (3.9 in. x 6.6 in. x 1.5 in.)
Probe Dimensions (Model 8760):
  Length:                           14.5 cm (5.81 in.) 
  Diameter:                         1.9 cm (0.75 in.) 
Probe Dimensions (Model 8762):
  Length:                           16.5 cm (6.50 in.) 
  Diameter:                         1.9 cm (0.75 in.) 
Weight (with batteries):
  Model 8732:                       0.34 kg (0.76 lbs) 
  Models 8760 and 8762:             0.53 kg (1.16 lbs) 
Power:                              Four AA-size batteries (included) or AC adapter 
                                    (included with Models 8760 and 8762)
Printer Interface:                  1200 baud rate, serial connection 


8732 8760 8762
CO2
CO    
Temperature  
Humidity  
% Outside Air  
Dew Point
Wet Bulb Temperature
Absolute Humidity
Humidity Ratio
Data Logging/Downloading
Statistics
Review Data  
Printer Output
NIST* Calibration Certificate
* U.S. National Institute of Standards and Technology.

All Specifications subject to change without notice.
1 At 25°C (77°F). Add uncertainty of ±0.36%/°C (±0.2%/°F) away from calibrated temperature.
2 For range of -5 to 95°C (23 to 203°F)
3 At 25°C (77°F). Add uncertainty of ±0.05% RH/°C (±0.03% RH/°F) away from calibrated temperature.

For technical comments or questions, e-mailemco@tsi.com.